C4H10 |
|
Performance
Measuring Range |
25 to 1400 ppm |
Number of Pump Strokes |
1 |
Correction Factor |
1 |
Sampling Time |
4 minutes per pump stroke |
Detecting Limit |
5 ppm (n=1) |
Colour Change |
Orange → Dark Green |
Reaction Principle |
C4H10 + Cr6++ H2SO4 → Cr3+ |
Coefficient of Variation |
10% (for 25 to 400 ppm), 5% (for 400 to 1400 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Organic solvents |
|
+ |
Dark green |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Isobutane |
Factor : 2.2 |
1 |
55 to 3080 ppm |
Pentane |
Factor : 1.2 |
1 |
30 to 1680 ppm |
Calibration gas generation |
Static gas dilution method |
TLV-TWA |
TLV-STEL |
Explosive range |
800 ppm |
- |
1.6 to 8.5% |