Methanol

NO.111LL
CH3OH
Performance
Measuring Range
2 to 20 ppm 20 to 56 ppm
Number of Pump Strokes
4 2
Correction Factor
1 2.8
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.2 ppm (n=4)
Colour Change Pale yellow → Pale bluish green
Reaction Principle CH3OH + Cr6+ + H2SO4 → Cr3+
Coefficient of Variation 15% (for 2 to 5 ppm), 10% (for 5 to 20 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Alcohols   + Pale bluish green
Aromatic hydrocarbons   + Pale bluish green
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
200 ppm 250 ppm 6 to 36%


Return to previous viewGo to top of page