Ethanol

NO.112
C2H5OH
Performance
Measuring Range
0.01 to 0.05% 0.05 to 2.5% 2.5 to 7.5%
Number of Pump Strokes
2 1 1/2
Correction Factor
0.2 1 3
Sampling Time 3 minutes per pump stroke
Detecting Limit 0.004% (n=2)
Colour Change Pink → Pale blue
Reaction Principle C2H5OH + Cr6+ + H2SO4 → Cr3+
Coefficient of Variation 15% (for 0.05 to 0.5 %), 10% (for 0.5 to 2.5 %)
Shelf Life 3 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Carbon monoxide   No No
Carbon dioxide   No No
Alcohols   + Pale blue
Calibration gas generation Static gas dilution method
TLV-TWA TLV-STEL Explosive range
1000 ppm - 3.3 to 19%


Return to previous viewGo to top of page