Ethanol

NO.112D
C2H5OH
Performance
Measuring Range 100 to 25000 ppm
Sampling Hours 1 to 10 hours
Detecting Limit 0.05 ppm (10 hours)
Colour Change Yellow → Brown
Reaction Principle C2H5OH + Cr6++ H2SO4 → Cr3+
Coefficient of Variation 15% (for 1000 to 3000 ppm-hr), 10% (for 3000 to 25000 ppm-hr)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Methanol,Esters   No No
Isopropyl alcohol 1/1 + Brown
Calibration gas generation Static gas dilution method
TLV-TWA TLV-STEL Explosive range
1000 ppm - 3.3 to 19%