Ethanol

NO.112L
C2H5OH
Performance
Measuring Range
50 to 100 ppm 100 to 2000 ppm
Number of Pump Strokes
2 1
Correction Factor
1/2 1
Sampling Time 4 minutes per pump stroke
Detecting Limit 5 ppm (n=2)
Colour Change Pink → Pale blue
Reaction Principle C2H5OH + Cr6+ + H2SO4 → Cr3+
Coefficient of Variation 15% (for 50 to 500 ppm), 10% (for 500 to 2000 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Alcohols   + Pale blue
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
1000 ppm - 3.3 to 19%