Isopropyl alcohol

NO.113
CH3CHiOHjCH3ii-C3H7OHj
Performance
Measuring Range
0.02 to 0.04% 0.04 to 2.5% 2.5 to 5.0%
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2
Sampling Time 3 minutes per pump stroke
Detecting Limit 0.004% (n=2)
Colour Change Pink → Pale Blue
Reaction Principle (CH3)2CHOH + Cr6+ + H2SO4 →Cr3+
Coefficient of Variation 15% (for 0.04 to 0.5 %), 10% (for 0.5 to 2.5 %)
Shelf Life 3 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Other alcohols   + Pale blue
Carbon dioxide   No No
Carbon monoxide   No No
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Propyl alcohol 1 1 0.04 to 2.5 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
200 ppm 400 ppm 2 to 12.7%