Benzene |
NO.121S |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 2 minutes per pump stroke | |||
Detecting Limit | 0.5 ppm (n=4) | |||
Colour Change | White → Dark green | |||
Reaction Principle | C6H6 + I2O5 + H2S2O7 → I2 | |||
Coefficient of Variation | 20% (for 5 to 40 ppm), 15% (for 40 to 120 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Hexane | 100 ppm | No | No |
Toluene | 200 ppm | No | No |
Xylene | 300 ppm | No | No |
Aromatic hydrocarbons other than benzene are trapped in the brown layer in the pretreatment tube. If the pretreatment reagent is entirely consumed (whole brown layer turns to dark brown), a higher reading will be given. |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
0.5 ppm | 2.5 ppm | 1.3 to 7.1% |