Benzene |
NO.121SL |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 2 minutes per pump stroke | ||
Detecting Limit | 0.25 ppm (n=5) | ||
Colour Change | White → Brownish gray | ||
Reaction Principle | C6H6 + I2O5 + H2S2O7 → I2 | ||
Coefficient of Variation | 20% (for 1 to 5 ppm), 15% (for 5 to 20 ppm) | ||
Shelf Life | 3 Years | ||
Corrections for temperature & humidity | Unnecessary | ||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Hexane | 50 ppm | No | No |
Toluene | 100 ppm | No | No |
Xylene | 150 ppm | No | No |
Aromatic hydrocarbons other than benzene are trapped in the brown layer in the pretreatment tube. If the pretreatment reagent is entirely consumed (whole brown layer turns to dark brown), a higher reading will be given. |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
0.5 ppm | 2.5 ppm | 1.3 to 7.1% |