Performance
Measuring Range |
2 to 500 ppm |
Sampling Hours |
1 to 10 hours |
Detecting Limit |
2 ppm (10 hours) |
Colour Change |
White → Brown |
Reaction Principle |
Toluene react with formaldehyde to form condensation polymer, which is brown in colour. |
Coefficient of Variation |
10% (for 20 to 500 ppm-hr) |
Shelf Life |
2 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Aromatic hydrocarbons |
|
+ |
Brown |
Alcohols, Ketones, Esters |
30ppm |
No |
No |
Other substance measurable with this detector tube
Substance |
Correction Factor |
Sampling Time |
Measuring Range |
Xylene |
Factor: 1.7 |
1 to 10 |
3.4 to 850 ppm |
Ethyl benzene |
Factor: 1.4 |
1 to 10 |
2.8 to 700 ppm |
Cumene |
Factor: 1.7 |
1 to 10 |
3.4 to 850 ppm |
Benzene |
Factor: 1.2 |
1 to 10 |
2.4 to 600 ppm |
Stylene |
Factor: 13 |
1 to 10 |
26 to 6500 ppm |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
20 ppm |
- |
1.4 to 6.7% |