C6H4iCH3j2 |
|
Performance
Measuring Range |
5 to10 ppm |
10 to 250 ppm |
250 to 625 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1.5 minute per pump stroke |
Detecting Limit |
1 ppm (n=2) |
Colour Change |
White → Brown |
Reaction Principle |
C6H4(CH3)2 + I2O5 + H2SO4 → I2 |
Coefficient of Variation |
10% (for 10 to 50 ppm), 5% (for 50 to 250 ppm) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Carbon Monoxide |
1000 ppm |
No(Two layers) |
Pale brown (Whole layer) |
Acetylene, Hexane |
2000 ppm |
No(Two layers) |
Pale brown (Whole layer) |
Toluene |
1/5 |
+ |
Brown |
Benzene |
1/5 |
+ |
Pale yellow |
Other substance measurable with this detector tube
Tube 123 Reading (n = 2) |
10 |
20 |
50 |
100 |
Trimethylbenzene Conc. (ppm) |
10 |
27 |
90 |
300 |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
100 ppm |
150 ppm |
1.0 to 7.0% |