Styrene |
NO.124 |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 30 seconds per pump stroke | |||
Detecting Limit | 1 ppm (n=2) | |||
Colour Change | White → Yellow | |||
Reaction Principle | C6H5CH:CH2+ H2S2O7 → Condensation Polymer | |||
Coefficient of Variation | 15% (for 20 to 100 ppm), 10% (for 100 to 500 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Alcohols | 0.5% | + (Bleaching) | No |
Esters | 0.5% | + (Bleaching) | No |
Ketones | 0.5% | + (Bleaching) | No |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
20 ppm | 40 ppm | 1.1 to 6.1% |