Styrene

NO.124L
C6H5CHFCH2
Performance
Measuring Range
2 to 25 ppm 25 to 100 ppm
Number of Pump Strokes
4 1
Correction Factor
1 4
Sampling Time 30 seconds per pump stroke
Detecting Limit 0.5 ppm (n=4)
Colour Change White → Yellow
Reaction Principle C6H5CH:CH2+ H2S2O7 → Condensation Polymer
Coefficient of Variation 10% (for 2 to 5 ppm), 5% (for 5 to 25 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Alcohols, Esters †10 times + (Bleaching) No
Aldehydes, Ketones †10 times + (Bleaching) No
Butadiene †5 times + (Bleaching) Dark brown
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Divinyl benzene 0.6 3 1 to 15 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
20 ppm 40 ppm 1.1 to 6.1%