Hydrogen cyanide

NO.12L
HCN
Performance
Measuring Range
0.36 to 1.25 ppm 1.25 to 2.5 ppm 2.5 to 60 ppm 60 to120 ppm
Number of Pump Strokes
5 2 1 1/2
Correction Factor
1/7 1/2 1 2
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.1 ppm (n=5)
Colour Change Yellow → Red
Reaction Principle 2HCN + HgCl2 → Hg(CN)2+ 2HCl
HCl + Base → Chloride product
Coefficient of Variation 10% (for 2.5 to 20 ppm), 5% (for 20 to 60 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Humidity correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
sulphur dioxide 20 ppm + Red
Hydrogen sulphide 5 ppm + Red
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Acetone cyanohydrine 1.15 1 2.88 to 69 ppm
Boron trichloride by scale 2 0.5 to 20 ppm
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
- C 4.7 ppm 5.6 to 40%