Hydrogen cyanide |
NO.12M |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1 minute per pump stroke | |||
Detecting Limit | 1 ppm (n=2) | |||
Colour Change | Yellow → Red | |||
Reaction Principle | 2HCN + HgCl2 → Hg(CN)2 + 2HCl HCl + Base → Chloride |
|||
Coefficient of Variation | 10% (for 50 to 200 ppm), 5% (for 200 to 800 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Hydrogen sulphide | 500 ppm | + | Red |
sulphur dioxide | 500 ppm | + | Red |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
- | C 4.7 ppm | 5.6 to 40% |