Vinylidene chloride |
NO.130L |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1 minute per pump stroke | |||
Detecting Limit | 0.1 ppm (n=2) | |||
Colour Change | Yellow → Reddish brown | |||
Reaction Principle | CH2:CCl2 + Cr6+ + H2SO4 → HCl HCl+ Base → Reddish brown product |
|||
Coefficient of Variation | 10% (for 1 to 4 ppm), 5% (for 4 to 14 ppm) | |||
Shelf Life | 2 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Substance | Concentration | Interference | Change colour by itself to |
Ethylene | 300 ppm | No | |
Vinyl chloride | 1/7 | + | Reddish brown |
Tetrachloroethylene | 4 times | + | Reddish brown |
Trichloroethylene | 1/1 | + | Reddish brown |
Toluene | 500 ppm | No | |
Benzene | 400 ppm | No |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
5 ppm | - | 5.6 to 11.4% |