CH2FCHC |
|
Performance
Measuring Range |
0.25 to 0.5 ppm |
0.5 to 1 ppm |
1 to 20 ppm |
20 to 54 ppm |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.05 ppm (n=4) |
Colour Change |
Yellow → Reddish brown |
Reaction Principle |
CH2:CHCl + Cr6+ + H2SO4 → HCl
HCl+ Base → Chlorides |
Coefficient of Variation |
10% (for 1 to 6 ppm), 5% (for 6 to 20 ppm) |
Shelf Life |
2 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Ethylene |
0.1% |
|
No |
Tetrachloroethylene |
3 times |
+ |
Reddish brown |
Trichloroethylene |
1/2 time |
+ |
Reddish brown |
Toluene, Xylene |
500 ppm |
|
No |
Benzene |
400 ppm |
|
No |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
1,3-Dichloropropane |
Factor: 0.5 |
2 |
0.5 to 10 ppm |
p-Ethyl benzylchloride |
Factor: 2.5 |
2 |
2.5 to 50 ppm |
Ethyl chloroformate |
Factor: 7 |
2 |
7 to 140 ppm |
2-Methyl allyl chloride |
Factor: 2.75 |
1 |
2.8 to 55 ppm |
Methyl chloroformate |
Factor: 58 |
5 |
58 to 1160 ppm |
Propylene dichloride |
Factor: 40 |
2 |
40 to 800 ppm |
1,2,4-Trichlorobenzene |
Factor: 0.65 |
4 |
0.65 to 13 ppm |
Calibration gas generation |
Permeation tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
1 ppm |
- |
3.6 to 23% |