Trichloroethylene

NO.132L
Cl2C:CHCl
Performance
Measuring Range
1 to 2 ppm 2 to 25 ppm 25 to 70 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2.8
Sampling Time 1 minute per pump stroke
Detecting Limit 0.4 ppm (n=2)
Colour Change Yellow → Purple
Reaction Principle Cl2C:CHCl + PbO2 + H2SO4 → HCl
HCl + Basic Compound → Chloride
Coefficient of Variation 10% (for 1 to 5 ppm), 5% (for 5 to 25 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Nitric oxide, Nitrogen oxides   No No
Hydrogen chloride, Chlorine, Bromine 200 ppm + Purple
Acetone   No No
Unsaturated halogenated hydrocarbons   + Purple
Aromatic hydrocarbons 100 ppm No
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Benzyl chloride 0.8 2 1.6 to 20 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
10 ppm 25 ppm -