Tetrachloroethylene |
NO.133L |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1.5 minutes per pump stroke | |||
Detecting Limit | 0.4 ppm (n=2) | |||
Colour Change | Yellow → Pink | |||
Reaction Principle | Cl2C:CCl2 + PbO2 + H2SO4 → HCl HCl + Base → Chloride |
|||
Coefficient of Variation | 10% (for 2 to 5 ppm), 5% (for 5 to 25 ppm) | |||
Shelf Life | 2 Years | |||
Corrections for temperature & humidity | Temperature correction is necessary | |||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Substance | Concentration | Interference | Change colour by itself to |
Nitric Oxide, Nitrogen dioxide | No | No | |
Hydrogen chloride, Chlorine, Bromine | + | Pink | |
Acetone | 200 ppm | No | No |
Unsaturated Halogenated HCs | + | Pink | |
Aromatic hydrocarbons | 100 ppm | No |
Substance | Correction Factor | No. of pump strokes | Measuring Range |
Pentachloroethane | 20 | 1 | 40 to 500 ppm |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
25 ppm | 100 ppm | - |