Carbon tetrachloride

NO.134
CCl4
Performance
Measuring Range
0.5 to 2.5 ppm 2.5 to 60 ppm
Number of Pump Strokes
2 to 5 1
Correction Factor
1/2 to 1/5 1
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 0.2 ppm (n=5)
Colour Change White → Yellow
Reaction Principle

CCl4 + I2O5 + H2S2O7 → COCl2
COCl2 + (CH3)2NC6 H4CHO→ (CH3)2NC6 H4Cl2
(CH3)2NC6 H4Cl2 + (C6H5)2NH → Yellow product

Coefficient of Variation 15% (for 2.5 to 20 ppm), 10% (for 20 to 60 ppm)
Shelf Life 1 Years
Corrections for temperature & humidity Unnecessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Hydrogen chloride 100 ppm + Yellow
Chlorine, Bromine 50 ppm + Yellow
Vinyl chloride, Methylene chloride   No No
Chloroform   No No
Methyl bromide 100 ppm + Yellow
Tetrachloroethylene   No No
1,1,1-Trichloroethane 100 ppm + Yellow
Trichloroethylene   No No
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Chloropicrin 1.0 1 25 to 60 %
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
5 ppm 10 ppm -