1,1,1-Trichloroethane |
NO.135 |
![]() |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 3 minutes per pump stroke | ||
Detecting Limit | 50 ppm (n=1) | ||
Colour Change | White → Reddish orange | ||
Reaction Principle | 1,1,1-Trichloroethane reacts with oxidizing agent to produce intermediate products then it react with detecting agent to produce reddish orange stain. | ||
Coefficient of Variation | 10% (for 100 to 200 ppm), 5% (for 200 to 500 ppm) | ||
Shelf Life | 3 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Chlorine, Bromine, Iodine | + | Reddish orange | |
Chloroform, Dichloromethane | + | Reddish orange | |
Carbon tetrachroride | No | No | |
Methyl bromide | + | Reddish orange | |
Trichloroethylene, Tetrachloroethylene | + | Reddish orange |
Substance | Correction Factor | No. of pump strokes | Measuring Range |
Chlorobromomethane | Factor: 0.22 | 1 | 22 to 110 ppm |
1,1-Dichloroethane | Factor: 0.9 | 1 | 90 to 4500 ppm |
1,2-Dichloroethane | Factor: 4.0 | 1 | 400 to 2000 ppm |
1,1,2-Trichloroethane | Tube 135 Reading (n=2) | 100 | 200 | 300 | 400 | 500 |
1,1,2-Trichloroethane (ppm) | 220 | 350 | 480 | 610 | 750 |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
350 ppm | 450 ppm | - |