1,1,1-Trichloroethane |
NO.135L |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 3 minutes per pump stroke | |||
Detecting Limit | 3 ppm (n=2) | |||
Colour Change | White → Pale Pink | |||
Reaction Principle | 1,1,1-Trichloroethane reacts with oxidizing agent to produce intermediate products then it reacts with detecting agent to produce pale pink stain. | |||
Coefficient of Variation | 10% (for 20 to 60 ppm), 5% (for 60 to 200 ppm) | |||
Shelf Life | 2 Years | |||
Corrections for temperature & humidity | Temperature correction is necessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Halogens | + | Pale pink | |
Nitrogen oxides | + | Pale pink | |
Saturated halogenated hydrocarbons | + | Pale pink |
Substance | Correction Factor | No. of pump strokes | Measuring Range |
1,2-Dichloroethane | 5.2 | 1 | 104 to 1040 ppm |
1,1,2,2-Tetrabromoethane | 0.046 | 4 | 0.92 to 9.2 ppm |
1,2,3-Trichloropropane | 1.8 | 4 | 36 to 360 ppm |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
350 ppm | 450 ppm | - |