Chloroform |
NO.137 |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 2 minutes per pump stroke | |||
Detecting Limit | 1 ppm (n=7) | |||
Colour Change | White → Orange | |||
Reaction Principle | CHCl3 + I2O5 + H2S2O7 → Cl2 Cl2 + o-Tolidine → Orange products |
|||
Coefficient of Variation | 15% (for 10 to 20 ppm), 10% (for 20 to 100 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Humidity correction is necessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Cl2, Br2, I2 | 1/20 time | + | Orange |
Unsaturated halogenated HCs | 1/20 time | + | Orange |
Saturated halogenated HCs | 1/10 time | + | Orange |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
10 ppm | - | - |