Methylene chloride

NO.138L
CH2Cl2
Performance
Measuring Range
4 to 10 ppm 10 to 60 ppm 60 to 150 ppm
Number of Pump Strokes
4 2 1
Correction Factor
0.4 1 2.5
Sampling Time 3 minutes per pump stroke
Detecting Limit 5 ppm (n=2)
Colour Change White → Pale Pink
Reaction Principle CH2:Cl2 + CrO3 + H2S2O7 → Cl2
Cl2+ 3,3,5,5-Tetramethylbenzidine → Holoquinone
Coefficient of Variation 15% (for 10 to 20 ppm), 10% (for 20 to 60 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Cl2, Br2, I2   + Pale pink
Unsaturated halogenated HCs 6 ppm + Pale pink
Saturated halogenated HCs 3 ppm + Pale pink
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
50 ppm - 15.5 to 66.9%