Ethyl acetate

NO.141L
CH3CO2C2H5
Performance
Measuring Range 20 to 800 ppm
Number of Pump Strokes 2
Correction Factor 1
Sampling Time 3 minutes per pump stroke
Detecting Limit 5 ppm (n=2)
Colour Change Yellow → Blackish brown → Light blue after few minute
Reaction Principle CH3CO2C2H5 + Cr6+ + H2SO4 → Cr3+
Coefficient of Variation 15% (for 25 to 200 ppm), 10% (for 200 to 800 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Alcohols   + Dark green
Ketones   + Dark green
Esters   + Dark green
Aromatic HCs   + Dark green
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Cymene Factor: 0.28 2 5.6 to 224 ppm
Diisopropyl benzene Factor: 0.5 1/2 10 to 400 ppm
Ethyl acrylate Factor: 0.4 2 8 to 320 ppm
2-Hexyl alcohol Factor: 3.0 3 60 to 2400 ppm
Isopropyl ether Factor: 0.9 2 18 to 720 ppm
Mesityl oxide Factor: 1.35 2 27 to 1080 ppm
Methyl acrylate Factor: 0.4 2 8 to 320 ppm
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
400 ppm - 2 to 11.5%