Butyl acetate |
NO.142L |
Measuring Range | 10 to 300 ppm |
Number of Pump Strokes | 2 |
Correction Factor | 1 |
Sampling Time | 4 minutes per pump stroke |
Detecting Limit | 2 ppm (n=2) |
Colour Change | Yellow → Dark brown → Pale blue after few minutes |
Reaction Principle | CH3CO2(CH2)3CH3 + Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation | 15% (for 10 to 100 ppm), 10% (for 100 to 300 ppm) |
Shelf Life | 2 Years |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Alcohols | + | Dark brown (few minutes later) → Pale blue | |
Ketones | + | ||
Esters | + | ||
Aromatic HCs | + |
Substance | Correction Factor | No. of pump strokes | Measuring Range |
Isobuty Acrylate | Factor: 0.26 | 2 | 2.6 to 78 ppm |
Butyl Acrylate | Factor: 0.7 | 2 | 7 to 210 ppm |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
150 ppm | 200 ppm | 1.7 to 7.6% |