Isopropyl acetate |
NO.146 |
Measuring Range | 10 to 500 ppm |
Number of Pump Strokes | 2 |
Correction Factor | 1 |
Sampling Time | 4 minutes per pump stroke |
Detecting Limit | 5 ppm (n=2) |
Colour Change | Yellow → Blackish brown → Pale blue after few minutes |
Reaction Principle | CH3CO2CH(CH3)2+ Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation | 15% (for 20 to 100 ppm), 10% (for 100 to 500 ppm) |
Shelf Life | 2 Years |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Alcohols | + | Blackish brown then turns to pale blue | |
Ketones | + | Blackish brown then turns to pale blue | |
Esters | + | Blackish brown then turns to pale blue |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
100 ppm | 200 ppm | 1.8 to 8% |