Amtyl acetate |
NO.147 |
Measuring Range | 10 to 200 ppm |
Number of Pump Strokes | 2 |
Correction Factor | 1 |
Sampling Time | 4 minutes per pump stroke |
Detecting Limit | 2 ppm (n=2) |
Colour Change | Yellow → Pale blue |
Reaction Principle | CH3CO2(CH2)4CH3 + Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation | 15% (for 10 to 50 ppm), 10% (for 50 to 200 ppm) |
Shelf Life | 2 Years |
Corrections for temperature & humidity | Temperature correction is necessary |
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Alcohols | + | Pale blue | |
Ketones | + | Pale blue | |
Esters | + | Pale blue |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
100 ppm | - | 1.1 to 7.5% |