Methyl metacrylate |
NO.149 |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 4 minutes per pump stroke | ||
Detecting Limit | 1 ppm (n=2) | ||
Colour Change | Yellow → Pale blue | ||
Reaction Principle | CH2:C(CH3)CO2CH3 + Cr6+ + H2SO4 → Cr3+ | ||
Coefficient of Variation | 10% (for 10 to 50 ppm), 5% (for 50 to 200 ppm) | ||
Shelf Life | 2 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Alcohols, Esters, Ketones | + | Pale blue |
Tube 102L Reading (n=2) | 10 | 20 | 30 | 50 | 100 | 150 | 200 |
Allyl isothiocyanate (%) | 5 | 10 | 15 | 30 | 70 | 130 | 200 |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
50 ppm | 100 ppm | 1.7 to 8.2% |