Hydrogen chloride(for Low Humidity)

NO.14R
HCl
Performance
Measuring Range
50 to 200 ppm 200 to 5000 ppm
Number of Pump Strokes
2 to 4 1
Correction Factor
1/2 to 1/4 1
Sampling Time 45 seconds per pump stroke
Detecting Limit 10 ppm (n=4)
Colour Change Purple → Pale pink
Reaction Principle Hydrogen chloride reacts with indicator to produce pale pink stain.
Coefficient of Variation 10% (for 200 to 800 ppm), 5% (for 800 to 2500 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Carbon monoxide   No No
Nitric oxide   No No
Chlorine 4ppm + White( 1ppm )
Carbon dioxide   No No
Nitrogen dioxide 1000ppm + White( 1000ppm )
Hydrogen sulphide 5% No
Sulphur dioxide 100ppm + Yellow
n-Hexane   No No
Calibration gas generation High pressure gas cylinder method
TLV-TWA TLV-STEL Explosive range
- C 2 ppm -