Acetone |
NO.151L |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 2 minutes per pump stroke | ||
Detecting Limit | 5 ppm (n=2) | ||
Colour Change | Yellow → Red | ||
Reaction Principle | CH3COCH3 + (NH2OH)3 • H3PO4 → H3PO4+ Base → H3PO4+ Base |
||
Coefficient of Variation | 15% (for 50 to 500 ppm), 10% (for 500 to 4000 ppm) | ||
Shelf Life | 2 Years | ||
Corrections for temperature & humidity | Temperature correction is necessary | ||
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Substance | Concentration | Interference | Change colour by itself to |
Acrolein, Acetaldehyde | 1/10 time | + | Red |
Aromatic hydrocarbons | No | No | |
Methyl isobutyl ketone | + | Red | |
Methyl ethyl ketone | + | Red |
Substance | Correction Factor | No. of pump strokes | Measuring Range |
Methyl ethyl ketone | Factor: 0.42 | 5 | 21 to 1680 ppm |
Propionaldehyde | Factor: 0.47 | 2 | 24 to 1880 ppm |
Calibration gas generation | Diffusion tube method |
TLV-TWA | TLV-STEL | Explosive range |
500 ppm | 750 ppm | 2.1 to 13% |