(CH3)2CHCH2COCH3 |
|
Performance
Measuring Range |
0.05 to 0.6% |
Number of Pump Strokes |
2 |
Correction Factor |
1 |
Sampling Time |
1.5 minutes per pump stroke |
Detecting Limit |
0.001% (n=2) |
Colour Change |
Orange →Dark green |
Reaction Principle |
(CH3)2CHCH2COCH3+ Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation |
10% (for 0.05 to 0.2 %), 5% (for 0.2 to 0.6 %) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
sulphur dioxide |
500 ppm |
+ |
Green |
Hydrogen sulphide |
500 ppm |
+ |
Dark green |
Alcohols, Esters,Ketones |
|
+ |
Dark green |
Propane |
0.2% |
+ |
Dark green (whole layer) |
Aromatic HCs |
500 ppm |
+ |
Dark green |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Styrene |
1.5 |
1 |
0.075 to 0.9% |
Calibration gas generation |
Static gas dilution method |
TLV-TWA |
TLV-STEL |
Explosive range |
50 ppm |
75 ppm |
1.2 to 8% |