Methyl isobutyl ketone

NO.153
(CH3)2CHCH2COCH3
Performance
Measuring Range 0.05 to 0.6%
Number of Pump Strokes 2
Correction Factor 1
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 0.001% (n=2)
Colour Change Orange →Dark green
Reaction Principle (CH3)2CHCH2COCH3+ Cr6+ + H2SO4 → Cr3+
Coefficient of Variation 10% (for 0.05 to 0.2 %), 5% (for 0.2 to 0.6 %)
Shelf Life 3 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
sulphur dioxide 500 ppm + Green
Hydrogen sulphide 500 ppm + Dark green
Alcohols, Esters,Ketones   + Dark green
Propane 0.2% + Dark green (whole layer)
Aromatic HCs 500 ppm + Dark green
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Styrene 1.5 1 0.075 to 0.9%
Calibration gas generation Static gas dilution method
TLV-TWA TLV-STEL Explosive range
50 ppm 75 ppm 1.2 to 8%