C6H10O |
|
Performance
Measuring Range |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
2 minutes per pump stroke |
Detecting Limit |
0.2 ppm (n=4) |
Colour Change |
Pale Yellow → Yellow |
Reaction Principle |
Cyclohexanone reacts with 2,4-dinitrophenylhydrazine to form dinitropnenylhydrazone to produce yellow in colour. |
Coefficient of Variation |
10% (for 2 to 10 ppm), 5% (for 10 to 30 ppm) |
Shelf Life |
2 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Furfural |
1.0 |
4 |
2 to 30 ppm |
Diacetone alcohol |
by scale |
2 |
2.5 to 100 ppm |
Isophorone |
1.0 |
8 |
2 to 30 ppm |
Calibration gas generation |
Diffusion tube method |
TLV-TWA |
TLV-STEL |
Explosive range |
25 ppm |
- |
1.1 to 9.4% |