C2H4O |
![](/files/images/icons/arrow_left_48.png) |
Performance
Measuring Range |
(0.05 ) to 3% |
Number of Pump Strokes |
1 |
Correction Factor |
1 |
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.02% (n = 1) |
Colour Change |
Orange → Green |
Reaction Principle |
C2H4O+ Cr6+ + H2SO4 → Cr3+ |
Coefficient of Variation |
10% (for 0.05 to 1.0 %), 5% (for 1.0 to 3.0 %) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Aromatic HCs |
|
+ |
Green |
Alcohols, Esters, Ethers |
|
+ |
Green |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
1,4-Dioxane |
Factor: 2.0 |
1 |
0.1 to 6% |
Propylene oxide |
Factor: 1.3 |
1 |
0.065 to 3.9% |
Calibration gas generation |
Static gas dilution method |
TLV-TWA |
TLV-STEL |
Explosive range |
1 ppm |
- |
3.6 to 100% |