Measuring Range |
10 to 100 mg/m3 |
Number of Pump Strokes |
2 |
Correction Factor |
1 |
Sampling Time |
2 minutes per pump stroke |
Detecting Limit |
2 mg/m3 (n=2) |
Colour Change |
Yellow → Reddish brown |
Reaction Principle |
HOCO2CH2OH + Oxidiser → 2HCHO
3HCHO + (NH2OH)3H3PO4 → H3PO4
H3PO4 + Base → Phosphate |
Coefficient of Variation |
15% (for 10 to 50 mg/m3),10% (for 50 to 150 mg/m3) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Temperature correction is necessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |