Ethylene

NO.172L
CH2¡§CH2
Performance
Measuring Range 0.2 to 50 ppm 50 to 100 ppm
Number of Pump Strokes 4 2
Correction Factor 1 2
Sampling Time 2 minutes per pump stroke
Detecting Limit 0.05 ppm (n=4)
Colour Change Pale yellow → Blue
Reaction Principle CH2:CH2 + (NH4)2MoO4 + PbSO4O → Molybdenum blue
Coefficient of Variation 10% (for 0.2 to 10 ppm), 5% (for 10 to 50 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ammonia, Hydrogen cyanide   + White
Carbon monoxide, Hydrogen   + Blue (whole layer)
Hydrogen chloride   + Pink
Hydrogen sulphide   + Black
Butadiene   + White
Butane, Pentane   + Blue (whole layer)
Butylene, Propylene 1/4 time + Blue
Calibration gas generation High pressure gas cylinder method
TLV-TWA TLV-STEL Explosive range
200ppm - 2.7 to 36%


Return to previous viewGo to top of page