1,3-Butadiene

NO.174L
CH2:CHCH:CH2
Performance
Measuring Range
2.5 to 5 ppm 5 to 100 ppm
Number of Pump Strokes
8 4
Correction Factor
1/2 1
Sampling Time 3 minutes per pump stroke
Detecting Limit 0.5 ppm (n=8)
Colour Change Pale yellow → White
Reaction Principle CH2:CHCH:CH2 + (NH4)2MoO4 + Pd2SO4 → White product
Coefficient of Variation 10% (for 5 to 20 ppm), 5% (for 20 to 100 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Acetylene, Hydrogen cyanide   + White
Carbon monoxide, Hydrogen   + Blue (whole layer)
Hydrogen chloride   + Pink
Hydrogen sulphide   + Black
Ethylene, Propylene, Butylene   + Blue
Butane, Pentane   + Blue (whole layer)
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
1,3-Pentadiene 8.5 4 42.5 to 850 ppm
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
2 ppm - 2.0 to 12%