1,3-Butadiene |
NO.174LL |
Measuring Range | 0.5 to 5 ppm |
Number of Pump Strokes | 1 |
Correction Factor | 1 |
Sampling Time | 3 minutes per pump stroke |
Detecting Limit | 0.1 ppm (n=1) |
Colour Change | Pink → Pale Yellow |
Reaction Principle | CH2:CHCH:CH2+ KMnO4 → Reaction Product |
Coefficient of Variation | 10% (for 0.05 to 1 ppm), 5% (for 2 to 5 ppm) |
Shelf Life | 3 Years |
Corrections for temperature & humidity | Unnecessary |
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. |
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | No | No | |
CO & CO2 | No | No | |
H2S, HCN | No | No | |
Acetylene | 50 ppm | + | White |
Olefins | + | Pale yellow |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
2 ppm | - | 2.0 to 12% |