CO |
![](/files/images/icons/arrow_left_48.png) |
Performance
Measuring Range |
0.05 to 0.1% |
0.1 to 2.0% |
2.0 to 4.0% |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1.5 minutes per pump stroke |
Detecting Limit |
0.01% (n=2) |
Colour Change |
White → Pale brown |
Reaction Principle |
5CO + I2O5 + H2S2O7 → I2 |
Coefficient of Variation |
10% (for 0.1 to 0.5 %), 5% (for 0.5 to 2 %) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Acetylene |
0.1% |
+ |
Brown |
Olefins |
2/3 time |
+ |
Pale brown |
Butane |
0.3% |
+ |
Pale brown (whole layer) |
Propane |
10% |
+ |
Pale brown (whole layer) |
Hexane |
0.1% |
+ |
Pale brown |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Calibration gas generation |
High pressure gas cylinder method |
TLV-TWA |
TLV-STEL |
Explosive range |
25 ppm |
- |
12.5 to 74% |