Carbon monoxide

NO.1M
CO
Performance
Measuring Range
0.05 to 0.1% 0.1 to 2.0% 2.0 to 4.0%
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 0.01% (n=2)
Colour Change White → Pale brown
Reaction Principle 5CO + I2O5 + H2S2O7 → I2
Coefficient of Variation 10% (for 0.1 to 0.5 %), 5% (for 0.5 to 2 %)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Acetylene 0.1% + Brown
Olefins 2/3 time + Pale brown
Butane 0.3% + Pale brown (whole layer)
Propane 10% + Pale brown (whole layer)
Hexane 0.1% + Pale brown
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Gasoline 1 1 0.1 to 2%
Calibration gas generation High pressure gas cylinder method
TLV-TWA TLV-STEL Explosive range
25 ppm - 12.5 to 74%


Return to previous viewGo to top of page