Chlorine Dioxide |
NO.23M |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1 minute per pump stroke | |||
Detecting Limit | 0.05 ppm (n=5) | |||
Colour Change | White → Pale pink | |||
Reaction Principle | Chlorine dioxide oxidises the indicator to produce Pale pink discolouration. | |||
Coefficient of Variation | 10% (for 0.05 to 0.2 ppm), 5% (for 0.2 to 0.6 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes in dark and cool place. |
Substance | Concentration | Interference | Change colour by itself to |
Nitric oxide | + | Pale pink | |
Ozone | + | Pale pink | |
Nitrogen dioxide | + | Pale pink | |
Chlorine | + | Pale pink | |
Bromine, Iodine | + | Pale pink |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
0.1 ppm | 0.3 ppm | - |