Carbon dioxide |
NO.2LC |
Measuring Range |
|
||
Number of Pump Strokes |
|
||
Correction Factor |
|
||
Sampling Time | 2 minutes per pump stroke | ||
Detecting Limit | 20 ppm (n=1) | ||
Colour Change | Pale Red → Yellow | ||
Reaction Principle | CO2+ 2KOH → K2CO3+ H2O | ||
Coefficient of Variation | 10% (for 100 to 600 ppm), 5% (for 700 to 2000 ppm) | ||
Shelf Life | 2 Years | ||
Corrections for temperature & humidity | Unnecessary | ||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | 1000 ppm | No | No |
Hydrogen chloride | 500 ppm | No | No |
Chlorine | 20 ppm | No | No |
HCN, H2S | 100 ppm | No | No |
sulphur dioxide | 25 ppm | No | No |
Nitorgen dioxide | 20 ppm | No | No |
Calibration gas generation | High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
5000 ppm | 30000 ppm | - |