NH3 |
|
Performance
Measuring Range |
0.2 to 1% |
1 to 16% |
16 to 32% |
|
Number of Pump Strokes |
|
Correction Factor |
|
Sampling Time |
1 minute per pump stroke |
Detecting Limit |
0.05% (n=5) |
Colour Change |
Purple → Yellow |
Reaction Principle |
3NH3 + H3PO4 → (NH4)3PO4 |
Coefficient of Variation |
10% (for 1 to 4 %), 5% (for 6 to 16 %) |
Shelf Life |
3 Years |
Corrections for temperature & humidity |
Unnecessary |
Store the tubes at cool and dark place. |
Possible coexisting substances and their interferences
(NOTE)
Substance |
Concentration |
Interference |
Change colour by itself to |
Hydrazine |
|
+ |
Yellow |
Amines other than aromatic amines |
|
+ |
Yellow |
Other substance measurable with this detector tube
Substance |
Correction Factor |
No. of pump strokes |
Measuring Range |
Dimethylamine |
1.2 |
1 |
1.2 to 19.2% |
Calibration gas generation |
High pressure gas cylinder method |
TLV-TWA |
TLV-STEL |
Explosive range |
25 ppm |
35 ppm |
15 to 28% |