Ammonia

NO.3HM
NH3
Performance
Measuring Range
0.05 to 1.6% 1.6 to 3.52%
Number of Pump Strokes
1 1/2
Correction Factor
1 2.2
Sampling Time 1 minute per pump stroke
Detecting Limit 0.01% (n=1)
Colour Change Pink → Yellow
Reaction Principle 2NH3 + H2SO4 → (NH4)2SO4
Coefficient of Variation 10% (for 0.05 to 0.4 %), 5% (for 0.4 to 1.6 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Hydrazine   + Yellow
Pyridine   + Yellow
Amines other than aromatic amines   + Yellow
Calibration gas generation High pressure gas cylinder method
TLV-TWA TLV-STEL Explosive range
25 ppm 35 ppm 15 to 28%


Return to previous viewGo to top of page