Ammonia |
NO.3L |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1 minute per pump stroke | |||
Detecting Limit | 0.2 ppm (n=2) | |||
Colour Change | Pink → Yellow | |||
Reaction Principle | 2NH3 + H2SO4 → (NH4)2SO4 | |||
Coefficient of Variation | 10% (for 1 to 10 ppm), 5% (for 10 to 30 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Carbon dioxide | 1% | No | |
Hydrazine | + | Yellow | |
Amines | 1/15 | + | Yellow ( 5 ppm) |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
25 ppm | 35 ppm | 15 to 28% |