Mercury vapour |
NO.40 |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1.5 minutes per pump stroke | |||
Detecting Limit | 0.01 mg/m3 (n=5) | |||
Colour Change | White → Pale Orange | |||
Reaction Principle | Hg + Cu2I2 → Cu2(HgI4) + 2Cu | |||
Coefficient of Variation | 10% (for 0.25 to 2 mg/m3), 5% (for 2 to 6 mg/m3) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Chlorine | + | Pale orange | |
Nitrogen dioxide | + | Pale orange | |
Hydrogen sulphide | + | Pale orange (whole layer) |
Calibration gas generation | Vaper pressure method |
TLV-TWA | TLV-STEL | Explosive range |
0.025mg/m3 | - | - |