Hydrogen sulphide

NO.4H
H2S
Performance
Measuring Range
10 to 100 ppm 100 to 2000 ppm 2000 to 4000 ppm
Number of Pump Strokes
2 to 10 1 1/2
Correction Factor
1/2 to 1/10 1 2
Sampling Time 30 seconds per pump stroke
Detecting Limit 2 ppm (n=10)
Colour Change White → Brown
Reaction Principle H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation 5% (for 100 to 2000 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ammonia   No No
Carbon monoxide,
Carbon dioxide
  No No
sulphur dioxide 1/1 + No
Mercaptans   No No
Organic substance   No No
Nitrogen dioxide 1/4 No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
10 ppm 15 ppm 4.0 to 44%


Return to previous viewGo to top of page