Hydrogen sulphide

NO.4HM
H2S
Performance
Measuring Range
25 to 50 ppm 50 to 800 ppm 800 to 1600 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2
Sampling Time 1 minute per pump stroke
Detecting Limit 5 ppm (n=2)
Colour Change White → Brown
Reaction Principle H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation 5% (for 50 to 800 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
@ @ @ @
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Nitrogen dioxide 1/4 time   No
sulphur dioxide 1/1 + No
Ammonia   No No
Carbon dioxide   No No
Carbon monoxide   No No
Mercaptans   No No
Organic gases & vapours   No No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
10 ppm 15 ppm 4.0 to 44%