Hydrogen sulphide |
NO.4HP |
Measuring Range |
|
|||
Number of Pump Strokes |
| |||
Correction Factor |
| |||
Sampling Time | 3 minutes per pump stroke | |||
Detecting Limit | 0.1% (n=2) | |||
Colour Change | Pale Blue → Blackish Brown | |||
Reaction Principle | H2S + CuSO4 → CuS + H2SO4 | |||
Coefficient of Variation | 10% (for 0.5 to 3 %), 5% (for 3 to 10 %) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Ammonia | Two layers (Dark brown & dark blue) |
Dark blue | |
sulphur dioxide | 2.5% | No | |
Mercaptans | 5 % | + | No |
Calibration gas generation | High pressure gas cylinder method |
TLV-TWA | TLV-STEL | Explosive range |
10 ppm | 15 ppm | 4.0 to 44% |