Hydrogen sulphide |
NO.4L |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 30 seconds per pump stroke | |||
Detecting Limit | 0.2 ppm (n=10) | |||
Colour Change | White → Brown | |||
Reaction Principle | H2S + Pb(CH3COO)2 → PbS + 2CH3COOH | |||
Coefficient of Variation | 5% (for 10 to 120 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Nitrogen dioxide | 1/4 time | No | |
Sulphur dioxide | 1/1 | + | No |
Mercaptans | No | No | |
Hydrogen chloride | Bleaching | No |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
10 ppm | 15 ppm | 4.0 to 44% |