Hydrogen sulphide |
NO.4LK |
Measuring Range |
|
|||
Number of Pump Strokes |
|
|||
Correction Factor |
|
|||
Sampling Time | 1 minute per pump stroke | |||
Detecting Limit | 0.25 ppm (n=2) | |||
Colour Change | White → Brown | |||
Reaction Principle | H2S + Pb(CH3COO)2 → PbS + 2CH3COOH | |||
Coefficient of Variation | 10% (for 2 to 6 ppm), 5% (for 6 to 20 ppm) | |||
Shelf Life | 3 Years | |||
Corrections for temperature & humidity | Unnecessary | |||
Store the tubes at cool and dark place. |
Substance | Concentration | Interference | Change colour by itself to |
Mercaptans | No | No | |
sulphur dioxide | 1/2 | + | No |
Nitrogen dioxide | 1/4 | No | |
Ozone | 10 ppm | Bleaching | No |
Calibration gas generation | Permeation tube method |
TLV-TWA | TLV-STEL | Explosive range |
10 ppm | 15 ppm | 4.0 to 44% |