Hydrogen sulphide

NO.4M
H2S
Performance
Measuring Range
12.5 to 25 ppm 25 to 250 ppm 250 to 500 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2
Sampling Time 30 seconds per pump stroke
Detecting Limit 2.5 ppm (n=2)
Colour Change White → Brown
Reaction Principle H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation 5% (for 25 to 250 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Nitrogen dioxide 1/4 timer No
sulphur dioxide 1/1 + No
Mercaptans   No No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
10 ppm 15 ppm 4.0 to 44%


Return to previous viewGo to top of page